This is version 16.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
Specifications
MA09 Detailed Manual
- on sample stick: x22642
- on VTI heat exchanger: c17844
- common, very rough calibration for all Allen-Bradley test sensors:
cplus70
Restrictions
Instrument | Field Limit | Comment |
TASP | 6 Tesla | with goniometer, goniometer motors removed |
TASP | 9 Tesla | without goniometer unit |
Rita-II | 4 Tesla | with goniometer, goniometer motors removed |
DMC | 1.4 Tesla | |
HRPT | ?? | force measurements to be redone |
FOCUS | 9 Tesla | when more than 1.5 m from guide shielding |
TriCS | 4 Tesla | |
Morpheus | 3.5 Tesla | with X-Y table, dist. to diaphragmas at least 250 mm |
Pending Issues
- The magnet has a cold leak, probably on the capillary, occuring
only occasionaly at low temperatures. The indium seal has been redone.
Because there is actually a high demand, as a workaround a diffusion
pump is connected on the main vacuum constantly. (2005-10-24)
- There seem to be a problem with the He level gauge. As a workaround, an
additional gauge is placed in the He filling port. (2005-10-19)
Attachments:
ma09b.jpg
taspMa9_1.jpg
ma09b.JPG
RCshielding.jpg
ma09a.jpg
RCmotor.jpg
ma09c.JPG
taspMa9_2.jpg
taspFilterMA09.png
This particular version was published on 11-Aug-2006 07:16:56 UTC by MarkusZolliker.
|
|