MA09


Specifications

More Information on intranet page (drawing): LdmWiki:MA09

ma09a.jpg ma09b.jpg

MA09 Detailed Manual

Temperature Sensors

Restrictions

Instrument Field Limit Comment
TASP 6 Tesla with goniometer, goniometer motors removed, distance to Be filter: > 20 cm
TASP 9 Tesla without goniometer unit, monochromator 2theta > 75, distance to Be filter[1]: > 22 cm
7.9 Tesla: 2theta > 55 ?
Rita-II 4 Tesla
9 Tesla
with goniometer, goniometer motors removed
without goniometer
DMC 1.4 Tesla
HRPT 4 Tesla without RC shielding [2]
FOCUS 9 Tesla magnetic rail and motor of radial collimator removed
distance magnet center to mono center 2 m, 2theta < 132 deg or magnet surface to guide surface > 0.9 m
TriCS 2 Tesla due to problems with omega-encoder (from the forces 4 T is possible)
Morpheus 3.5 Tesla with X-Y table, dist. to diaphragmas at least 250 mm, omega motor not in use
Morpheus 2 Tesla with omega motor
[1] on TASP: MA09/taspFilterMA09.png

[2] on HRPT: MA09/RCshielding.jpgMA09/RCmotor.jpg
The RC shielding (left picture) has to be removed. One of the pieces making trouble is probably the RC motor (right picture).

On all other Instruments, MA09 can not yet be used.

Pending Issues

Solved issues

TASP adapter TASP adapter