More Information on intranet page (drawing): LdmWiki:MA09
Instrument | Field Limit | Comment |
---|---|---|
TASP | 6 Tesla | with goniometer, goniometer motors removed, distance to Be filter: > 20 cm |
TASP | 9 Tesla | without goniometer unit, monochromator 2theta > 75, distance to Be filter[1]: > 22 cm 7.9 Tesla: 2theta > 55 ? |
Rita-II | 4 Tesla 9 Tesla | with goniometer, goniometer motors removed without goniometer |
DMC | 1.4 Tesla | |
HRPT | 4 Tesla | without RC shielding [2] |
FOCUS | 9 Tesla | magnetic rail and motor of radial collimator removed distance magnet center to mono center 2 m, 2theta < 132 deg or magnet surface to guide surface > 0.9 m |
TriCS | 2 Tesla | due to problems with omega-encoder (from the forces 4 T is possible) |
Morpheus | 3.5 Tesla | with X-Y table, dist. to diaphragmas at least 250 mm, omega motor not in use |
Morpheus | 2 Tesla | with omega motor |
[2] on HRPT:
The RC shielding (left picture) has to be removed. One of the pieces making trouble is probably the RC motor (right picture).
On all other Instruments, MA09 can not yet be used.