DIL Sample mounting

See also Calculator for Sample Mounting in DIL

used in VARIOXMA15 MA11 MA02MA09MA6MA7MA10MB11VarioxB
max. sample diameter* (incl. holder) [mm] d40 14 14 21 40 21 32 40 77 77
sample space diameter [mm] 42.817 17 23.1 42.8 23.4 35.8 42.8 85.8 (77.4) 85.8 (77.4)
distance from DIL2 mixing chamber to beam center [mm] l83+296 296 342+-5 272 295+-5 n.a. ?
distance from DIL3/DIL4 mixing chamber to beam center [mm]l61+296 296 342+-5 250 295+-5 350 ?
distance from DIL5 mixing chamber to beam center [mm] l 210+-10 210+-10
allowed size below beam center [mm]b 32 32 60 24** 65 59 32 150 150
(*) The maximum allowed size depends on the distance to the next spacer, which is larger when using the weak link. Please contact Markus Zolliker for designing sample holders slightly larger than the recommended max. sample diameter.
(**) Slightly higher values might be acceptable on request.

On the mixing chamber as well as at the end of the sample holder/weak link there is an M6 hole, 6 mm deep. The weak link has a length of 30 mm.

For some experiments, it might be possible to fit 2 samples in the DIL insert, making possible to measure 2 samples in sequence without removing the insert (saving one working day). For MA11 and MA15, the sample holder should look like the following picture.

DilSampleMounting/Dil2Samples.png

Drawings of the DIL cans can be found at the intranet http://lns00/ldmwiki/Wiki.jsp?page=DIL


Attachments:  Dil2Samples.png 