Oxford Instruments Kelvinox Dilution Insert DIL2

DIL2 was aquired in 2006, mainly financed by the "SNF Förderungsprofessur Michel Kenzelmann". It is equipped with a weak link, allowing access to the range 1...2 K without removing the mixture.

Sample Mounting

DIL1/dilGeometry.gif

Problems with DIL2

  • The weak link was too weak, and there was no spacer on it. We have shortened the weak section and put a spacer above it.

  • The gauges G1 and G2 do not work reliably. As a workaround, the (earlier unused) G3 gauge has now a connection to the beam dump.

Preparing DIL

Condense Mixture

Remove Mixture

Warming up

Sensor Calibration

The calibration curve in the FemtoPower system from Oxford Instruments is generic. For more precise values use the table below.
DisplayMix.Ch.
0.0400.042
0.0600.063
0.0800.084
0.1000.105
0.1500.156
0.2000.206
0.3000.306
0.4000.404
0.6000.600
0.8040.800
1.01 1.00
1.52 1.50
2.03 2.00
3.04 3.00
4.02 4.00
5.00 5.00


Attachments:  dil2Open.jpg  GasHandling.jpg  dil2.jpg 