Removing the Mixture
It is highly recommended to use the automatic scripts in SEA.
You might choose between:
- remove_only: remove the mixture and stay at 10 K
- remove_and_warmup: remove and warm up to 80 K for removing the stick
- remove_and_condense
- remove_and_sorbump
Manual procedure:
- close valve 13A, open valves 9 and 3
- set the mixing chamber heater and the still heater to 19 mW
- set the VTI temperature to 4 K
- the VTI set temperature can be set to 10 K, as soon as the dump
pressure is beyond 90 % of its original value (630 mbar on DIL1,
500 mbar on DIL2)
- wait until the pressure on P1 drops below 0.1 mbar, then close
all valves, and stop the 3He pump.