Preparing the Kelvinox Dilution and Gas Handling System

  • pump the traps (heat out each for 30 min, but pump longer)
  • pump out the condenser and pumping line, on the gas handling system and on the stick (all volumes might be connected)

After some hours (or one night)

  • make the CAF seal (red silicon paste) with clean fingers
  • put the tube, turning it a little on the seal and start pumping with the leak detector.
  • leak check
  • put the stick into the Variox or magnet VTI.
  • insert about 50 mbar 4He into the IVC
  • take the special assembly to connect the pumping line and the condenser line with the stick
  • pump out the adapter piece
  • connect a turbo pump to the IVC, start te pump, but do not yet open the IVC valve.

The following steps can be done automatically in SEA, under "main diltuion control"

  • ("cool") cool down the heat exchanger of the variox or magnet
  • When the stick is cold (< 4K, ins SEA "readyforpump") pump the IVC by clicking on "sorbpump" and opening the IVC valve, which is already connected to a running turbo pump. This heats up the VTI to 5 K, which might take a while. After 20 min. with sorb temperature set to 40 K, then 15 min without sorb heater, the "condense" step starts automatically.