Specifications
MA09 Detailed Manual
- on sample stick: x22642
- on VTI heat exchanger: c17844
- common, very rough calibration for all Allen-Bradley test sensors:
cplus70
Restrictions
Instrument | Field Limit | Comment |
TASP | 6 Tesla | with goniometer, goniometer motors removed |
TASP | 9 Tesla | without goniometer unit (adapter piece should be available in Sept 2005) |
Rita-II | 4 Tesla ? | with goniometer, goniometer motors removed (to be remeasured) |
DMC | 1.4 Tesla | |
HRPT | ?? | force measurements to be redone |
FOCUS | 9 Tesla | when more than 1.5 m from guide shielding |
TriCS | ?? | force measurements to be redone |
Pending Issues
- The magnet has a cold leak, probably on the capillary, occuring
only occasionaly at low temperatures. The indium seal has been redone.
Because there is actually a high demand, as a workaround a diffusion
pump is connected on the main vacuum constantly. (2005-10-24)
- There seem to be a problem with the He level gauge. As a workaround, an
additional gauge is placed in the He filling port. (2005-10-19)