Oxford Instruments Kelvinox Dilution Insert DIL2

DIL2 was aquired in 2006, mainly financed by the "SNF Förderungsprofessur Michel Kenzelmann". It is equipped with a weak link, allowing access to the range 1...2 K without removing the mixture.

Problems with DIL2

  • The weak link is too weak. For the first experiments we will use a simple copper sample holder, and use the sensor on the mixing chamber.

DIL2 Sample mounting

used inMA15MA11MA02MA09
sample space diameter [mm]17.317.323.645
max. sample diameter [mm]12121835
distance from thread to beam center [mm]
adjustable by about +- 10 mm
405197194
allowed size below beam center [mm]64306024
On the mixing chamber, there is a bolt with an M5 thread (on the sample container a 7 mm deep hole with an M5 thread is needed).

Preparing DIL

Condense Mixture

Remove Mixture

Warming up

Sensor Calibration

The calibration curve in the FemtoPower system from Oxford Instruments is generic. For more precise values use the table below.

Mix.Ch.DisplaySample
0.0430.0400.036
0.0630.0600.055
0.0840.0800.075
0.1040.1000.094
0.1550.1500.143
0.2050.2000.191
0.3060.3000.289
0.4070.4000.387
0.6090.6000.581
0.8110.8000.777
1.0131.0000.972
1.5211.5001.461
2.0292.0001.951
3.0503.0002.938
4.0754.0003.929
5.1055.0004.925


Attachments:  dil2Open.jpg  GasHandling.jpg  dil2.jpg 