Oxford Instruments Kelvinox Dilution Insert DIL2
DIL2 was aquired in 2006, mainly financed by the
"SNF Förderungsprofessur Michel Kenzelmann".
It is equipped with a weak link, allowing access to the
range 1...2 K without removing the mixture.
DIL2 Sample mounting
used in | | MA15 | MA11 | MA02 | MA09 |
sample space diameter [mm] | | 17.3 | 17.3 | 23.6 | 45 |
max. sample diameter [mm] | d | 16 | 16 | 18 | 35 |
distance from M6 thread to beam center [mm] adjustable by about +- 10 mm | a | 51 | 51 | 97 | 194 |
allowed size below beam center [mm] | b | 35 | 35 | 60 | 24 |
On the mixing chamber as well as at the end of the
sample holder there is an M6 hole.
Problems with DIL2
- The weak link was too weak, and there was no spacer on it.
We have shortened the weak section and put a spacer above
it.
- The gauges G1 and G2 do not work reliably. As a workaround,
the (earlier unused) G3 gauge has now a connection to the
beam dump.
Preparing DIL
Condense Mixture
Remove Mixture
Warming up
Sensor Calibration
The calibration curve in the FemtoPower system from Oxford Instruments
is generic. For more precise values use the table below.
Display | Mix.Ch. |
0.040 | 0.042 |
0.060 | 0.063 |
0.080 | 0.084 |
0.100 | 0.105 |
0.150 | 0.156 |
0.200 | 0.206 |
0.300 | 0.306 |
0.400 | 0.404 |
0.600 | 0.600 |
0.804 | 0.800 |
1.01 | 1.00 |
1.52 | 1.50 |
2.03 | 2.00 |
3.04 | 3.00 |
4.02 | 4.00 |
5.00 | 5.00 |