Oxford Instruments Kelvinox Dilution Insert DIL2

DIL2 was aquired in 2006, mainly financed by the "SNF Förderungsprofessur Michel Kenzelmann". It is equipped with a weak link, allowing access to the range 1...2 K without removing the mixture.

Problems with DIL2

  • The weak link is too weak. For the first experiments we will use a simple copper sample holder, and use the sensor on the mixing chamber.

DIL2 Sample mounting

used inMA15MA11MA02MA09
sample space diameter [mm]17.317.323.645
max. sample diameter [mm]16161835
distance from M6 thread to beam center [mm]
adjustable by about +- 10 mm
515197194
allowed size below beam center [mm]35356024
On the mixing chamber as well as at the end of the sample holder there is an M6 hole.

Preparing DIL

Condense Mixture

Remove Mixture

Warming up

Sensor Calibration

The calibration curve in the FemtoPower system from Oxford Instruments is generic. For more precise values use the table below.
Mix.Ch.DisplaySample
0.0430.0400.036
0.0630.0600.055
0.0840.0800.075
0.1040.1000.094
0.1550.1500.143
0.2050.2000.191
0.3060.3000.289
0.4070.4000.387
0.6090.6000.581
0.8110.8000.777
1.0131.0000.972
1.5211.5001.461
2.0292.0001.951
3.0503.0002.938
4.0754.0003.929
5.1055.0004.925