Lines 24-26 were replaced by lines 24-30
- * The magnet has suffered of some cold leak, occuring only occasionaly
- at low temperatures. An indium seal on the capillary has been redone
- --> __solved__.
+ * The magnet has a cold leak, probably on the capillary, occuring
+ only occasionaly at low temperatures. The indium seal has been redone.
+ Because there is actually a high demand, as a workaround a diffusion
+ pump is connected on the main vacuum constantly. (2005-10-24)
+
+ * There seem to be a problem with the He level gauge. As a workaround, an
+ additional gauge is placed in the He filling port. (2005-10-19)

Back to MA09, or to the Page History.