Cryomagnets

Please read also the rules concerning superconducting magnet experiments
and the Introduction on cryogenic handling
Our standard thread on all systems is M6 female.

NameDescriptionFieldTemp.d
(without
dilution)
a
[mm]
b
[mm]
l
[mm]
g
[mm]
h
[mm]
MA15Oxford Instr. Cryomagnet vert. 14.9 T 1.6...300 K
>80 mK[1]
20 14...90 139 1294 (1454)49.51078(+2)
MA11Oxford Instr. SANS Cryomagnethor. 11 T 1.8...300 K
>80 mK[1]
22 75...100[3]541302 (1462)62 1080(+2)
MA09Oxford Instr. Cryomagnet vert. 9 T 1.8...300 K
>80 mK[1]
48 93...137[3]581247 55
MA7 Oxford Instr. Cryomagnet hor. 6.8 T 1.8...300 K
>80 mK[1]
48 35...90 69 1357 (1517)49.5
MA6 Oxford Instr. Cryomagnet vert. 6 T 1.8...300 K
>80 mK[1]
30 40...90 80 1309 (1469*)49.51078(+2)
MA02Oxford Instr. Cryomagnet hor. 1.8 T 1.8...150 K[4]
>100 mK[1]
28 0...56 93 1245 (1405) 55 968(+2)
MA10Oxford Instr. Cryomagnet vert. 10 T 1.8...300 K
>80 mK[1]
48 1115 55

where
d: sample space diameter (reduce by 2-4 mm for max. sample diameter)
a: distance from thread to beam center
b: allowed size below beam center
l: length from top flange to the center of the sample (when used with a rotation unit, add 160)
g: diameter of upper part of sample stick space
h: length of upper (larger) part of sample stick space (+2mm when warm)
*: MA6 with rotation, measured on TriCS (geometrical: 1464)
The allowed sample diameter in combination with a dilution can be found here

[1] in combination with dilution sticks
[3] as we have adapter pieces, it is recommended to make the sample holder with a distance from thread to the sample center between 37 and 50 mm
[4] for a short time 300 K is possible, but not recommended due to very high liquid He consumption

Strayfields

SampleMountingInstructions

Magnetic Forces

Electromagnets

InstrumentsDescriptionField
TASP Guide field magnet vert. 20 mT
hor. 2 mT
SANS-I, SANS-II, AMORElectromagnet hor. < 1.5 T