This is version 42.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
The sample mounting thread on all sticks is M6.
name | description | T-range | d | a | b | l | t | w | h |
ILL1 | 50 mm Orange cryostat | 1.5...320 K | 50 | 15...49 | 44 | 916 | +-29 |
ILL2 | 70 mm Orange cryostat | 1.5...320 K | 68 | 13...89 | 70 | 912 | +-28 |
ILL3 | 70 mm Cryo-Furnace | 2...450 K | 68 | 19...97 | 71 | 944 | |
ILL4 | 70 mm Orange cryostat (FOCUS) | 1.5...320 K | <60 | 29...117 | 60 | 944 | |
ILL5 | 100 mm Orange cryostat | 1.5...320 K | 100 | 31...58 | 68 | 1128 | +-25 | 1199 | 1012 |
ORI1 | 50 mm Orange cryostat | 1.5...320 K | 50 | 37...117 | 56 | 956 | +-41 |
ORI2 | 50 mm Orange cryostat | 1.5...320 K | 50 | 31...89 | 56 | 957 | +-41 |
ORI3 | 100 mm Orange cryostat | 1.5...320 K | 100 | 21...65 | 66 | 1139 | | 1206 | 1016 |
ORI4 | 100 mm Orange cryostat (HRPT) | 1.5...320 K | 100 | 14...55 | 70 | 1128 | |
ORI6 | 100 mm Orange cryostat (EIGER) | 1.5...320 K | 100 | 60...80 | 110 | 1149 | | 1257 | 1016 | 1069 |
where
d: sample space diameter (subtract 2 mm for tolerance)
a: distance from thread to beam center
b: allowed size below beam center
l: length of the sample stick from the top flange to the center of the sample
see also sample mounting instructions and adapter pieces
t: vertical opening angle more info
w: warm total sample height
h: warm distance to heat exchanger edge
Additonal Information on Orange Cryostats
Obsolete information
He flow 4-circle cryostat (no longer in use): HEF4C
Attachments:
RisoeOrangeCryostatManual.pdf
ILL1a.jpg
ORIdimensions.png
This particular version was published on 22-Jan-2018 13:23:53 UTC by 129.129.156.86.
|