name | description | T-range | d | a | b | L | t | w | h |
---|---|---|---|---|---|---|---|---|---|
ILL1 | 50 mm Orange cryostat | 1.5...320 K | 50 | 15...49 | 44 | 916 | +-29 | ||
ILL2 | 70 mm Orange cryostat | 1.5...320 K | 68 | 13...89 | 70 | 912 | +-28 | ||
ILL3 | 70 mm Cryo-Furnace | 2...450 K | 68 | 19...97 | 71 | 944 | |||
ILL4 | 70 mm Orange cryostat (FOCUS) | 1.5...320 K | <60 | 29...117 | 60 | 944 | |||
ILL5 | 100 mm Orange cryostat | 1.5...320 K | 100 | 31...58 | 68 | 1128 | +-25 | 1199 | 1012 |
ORI1 | 50 mm Orange cryostat | 1.5...320 K | 50 | 37...117 | 56 | 956 | +-41 | ||
ORI2 | 50 mm Orange cryostat | 1.5...320 K | 50 | 31...89 | 56 | 957 | +-41 | ||
ORI3 | 100 mm Orange cryostat | 1.5...320 K | 100 | 21...65 | 66 | 1139 | 1206 | 1016 | |
ORI4 | 100 mm Orange cryostat (HRPT) | 1.5...320 K | 100 | 14...55 | 70 | 1128 | |||
ORI6 | 100 mm Orange cryostat (EIGER) | 1.5...320 K | 100 | 60...80 | 110 | 1149 | 1257 | 1016 | |
VARIOX | 50 mm cryostat for DIL | 1.5...320 K | 50 | ? | ? | 1068 | |||
VarioxB | 100 mm Variox Cryostat for P-cells | 1.8..300K >50mK[1] | 100 | 50...150 | 150 | 1551 (permanent stick rot.) | +-15 | 1551 |
where
d: sample space diameter (subtract 2 mm for tolerance)
a: distance from thread to beam center
b: allowed size below beam center
L: length of the sample stick from the top flange to the center of the sample
see also sample mounting instructions and adapter pieces. Please be aware that the values given are to measure at RT, a 2-3 mm shrinking at low T is already considered
t: vertical opening angle more info
w: warm total sample height
h: warm distance to heat exchanger edge
He flow 4-circle cryostat (no longer in use): HEF4C