| At line 7 added 3 lines. |
| + A rotatable sample stick has been made at PSI using a similar design at the |
| + ILL. This allows to leave the magnet in a fixed omega position and to |
| + turn the sample orientation within the magnet. |
| At line 8 added 3 lines. |
| + At present, MA15 can only be used on RITA2. However, some |
| + [Restrictions|MA15RitaRestrictions] have to be obeyed. |
| + |
| Line 13 was replaced by lines 19-27 |
| - !Actual Problems |
| + !Hanging Issues |
| + |
| + * The automatic needle vavle control does not work properly. It seems that this |
| + is a mechanical problem of the cold valve. The Magnet has to be sent back to |
| + Oxford Instruments. |
| + |
| + * The rotatable sample stick has a relatively bad thermal contact. Therefore the |
| + minimal sample temperature is increased. It is not sure if putting more |
| + exchange gas can elimate this problem. |
| At line 14 added 6 lines. |
| + * In order to be used possibly on other instruments, the instruments have to be |
| + checked for magnetic materials around the sample table. After some force |
| + tests, and after eventual modifications of the instrument, the conditions |
| + for a safe operations have to be defined. The next candidate for these |
| + tests is TASP. |
| + |