Difference between version 43 and version 4:
Lines 2-6 were replaced by lines 2-7
- partly by PSI, Risø, MaNEP and SNF. 14.9 Tesla is the maximum
- field achievable with nowadays technology for a split pair coil (which
- is needed for wide angle access). This limit could in principle be
- increased to 17.5 by means of a dysprosium insert for small samples.
- This option is not (yet) available at SINQ.
+ partly by
+ * [PSI|http://www.psi.ch]
+ * the [Carlsberg Foundation|http://www.carlsbergfondet.dk/]
+ and the Danish National Science Foundation (danish: [FNU|http://www.forsk.dk])
+ * [MaNEP|http://www.manep.ch]
+ * [SNF|http://www.snf.ch]
At line 7 added 1 line.
+ [ma15d.jpg]
Line 9 was replaced by lines 11-14
- [Specifications|MA15Specifications]
+ 14.9 Tesla is the maximum field achievable with nowadays technology
+ for a split pair coil (which is needed for wide angle access). This
+ limit could in principle be increased to 17.5 by means of a dysprosium
+ insert for small samples. This option is not (yet) available at SINQ.
At line 10 added 22 lines.
+ A rotatable sample stick has been made at PSI using a similar design at the
+ ILL. This allows to leave the magnet in a fixed omega position and to
+ turn the sample orientation within the magnet.
+
+ !Restrictions
+
+ __after several quenches, operation above 12.5 without lambda stage is not recommended [see|http://samenv:8080/sample_environment/61]__
+
+ Due to force limits, restrictions are defined for the use of MA15.
+ ||Instrument ||Allowed Max. Field [[Tesla]
+ |RITA-2 |11.5 ... 14.9, see [MA15 Rita Restrictions]
+ |TASP |14.9, see [MA15 TASP Restrictions]
+ |EIGER |13, see [MA15 Eiger Restrictions]
+ |ZEBRA |14.9 for the longer distance as defined for MA15
+ On all other instruments, MA15 can not yet be used.
+
+ [Specifications / Ramp Limits|MA15Specifications]
+
+ More information on intranet page (Drawing): [LdmWiki:MA15]
+
+ [Omega Motor Change on Rita2]
+
Line 13 was replaced by lines 40-59
- !Actual Problems
+ !Pending Issues
+
+ * The automatic needle valve does not always work properly.
+ A mechanical problem was solved by Oxford Instruments.
+ It is still a problem that sometimes the needle valve bar
+ gets frozen. This problem might be avoided when keeping warm
+ the top of the needle valve during fill. However, at high
+ fields, the air dryer might not work!
+
+ * The nitrogen level gauge does not work. This has to be fixed, but
+ it seems that it may appear often.
+
+ * During He-fill, LN2 has to be filled again in order to avoid
+ air sucking into the LN2 dewar.
+
+ * In order to be used possibly on other instruments, the instruments have to be
+ checked for magnetic materials around the sample table. After some force
+ tests, and after eventual modifications of the instrument, the conditions
+ for a safe operations have to be defined. The next candidate for these
+ tests is TASP.
At line 14 added 3 lines.
+ [ma15b.jpg]
+ [ma15a.jpg]
+ [ma15c.jpg]

Back to MA15, or to the Page History.