Lines 2-6 were replaced by lines 2-7 |
- partly by PSI, Risø, MaNEP and SNF. 14.9 Tesla is the maximum |
- field achievable with nowadays technology for a split pair coil (which |
- is needed for wide angle access). This limit could in principle be |
- increased to 17.5 by means of a dysprosium insert for small samples. |
- This option is not (yet) available at SINQ. |
+ partly by |
+ * [PSI|http://www.psi.ch] |
+ * the [Carlsberg Foundation|http://www.carlsbergfondet.dk/] |
+ and the Danish National Science Foundation (danish: [FNU|http://www.forsk.dk]) |
+ * [MaNEP|http://www.manep.ch] |
+ * [SNF|http://www.snf.ch] |
At line 7 added 1 line. |
+ [ma15d.jpg] |
Line 9 was replaced by lines 11-14 |
- [Specifications|MA15Specifications] |
+ 14.9 Tesla is the maximum field achievable with nowadays technology |
+ for a split pair coil (which is needed for wide angle access). This |
+ limit could in principle be increased to 17.5 by means of a dysprosium |
+ insert for small samples. This option is not (yet) available at SINQ. |
At line 10 added 22 lines. |
+ A rotatable sample stick has been made at PSI using a similar design at the |
+ ILL. This allows to leave the magnet in a fixed omega position and to |
+ turn the sample orientation within the magnet. |
+ |
+ !Restrictions |
+ |
+ __after several quenches, operation above 12.5 without lambda stage is not recommended [see|http://samenv:8080/sample_environment/61]__ |
+ |
+ Due to force limits, restrictions are defined for the use of MA15. |
+ ||Instrument ||Allowed Max. Field [[Tesla] |
+ |RITA-2 |11.5 ... 14.9, see [MA15 Rita Restrictions] |
+ |TASP |14.9, see [MA15 TASP Restrictions] |
+ |EIGER |13, see [MA15 Eiger Restrictions] |
+ |ZEBRA |14.9 for the longer distance as defined for MA15 |
+ On all other instruments, MA15 can not yet be used. |
+ |
+ [Specifications / Ramp Limits|MA15Specifications] |
+ |
+ More information on intranet page (Drawing): [LdmWiki:MA15] |
+ |
+ [Omega Motor Change on Rita2] |
+ |
Line 13 was replaced by lines 40-59 |
- !Actual Problems |
+ !Pending Issues |
+ |
+ * The automatic needle valve does not always work properly. |
+ A mechanical problem was solved by Oxford Instruments. |
+ It is still a problem that sometimes the needle valve bar |
+ gets frozen. This problem might be avoided when keeping warm |
+ the top of the needle valve during fill. However, at high |
+ fields, the air dryer might not work! |
+ |
+ * The nitrogen level gauge does not work. This has to be fixed, but |
+ it seems that it may appear often. |
+ |
+ * During He-fill, LN2 has to be filled again in order to avoid |
+ air sucking into the LN2 dewar. |
+ |
+ * In order to be used possibly on other instruments, the instruments have to be |
+ checked for magnetic materials around the sample table. After some force |
+ tests, and after eventual modifications of the instrument, the conditions |
+ for a safe operations have to be defined. The next candidate for these |
+ tests is TASP. |
At line 14 added 3 lines. |
+ [ma15b.jpg] |
+ [ma15a.jpg] |
+ [ma15c.jpg] |