Lines 21-22 were replaced by lines 21-29 |
- Due to force limits, restrictions are defined for the use of MA15.\\ |
- See [RITA|MA15RitaRestrictions], [TASP|MA15TASPRestrictions] or [EIGER|MA15EigerRestrictions].\\ |
+ |
+ __after several quenches, operation above 12.5 without lambda stage is not recommended [see|http://samenv:8080/sample_environment/61]__ |
+ |
+ Due to force limits, restrictions are defined for the use of MA15. |
+ ||Instrument ||Allowed Max. Field [[Tesla] |
+ |RITA-2 |11.5 ... 14.9, see [MA15 Rita Restrictions] |
+ |TASP |14.9, see [MA15 TASP Restrictions] |
+ |EIGER |13, see [MA15 Eiger Restrictions] |
+ |ZEBRA |14.9 for the longer distance as defined for MA15 |
Line 25 was replaced by lines 32-34 |
- [Specifications|MA15Specifications] |
+ [Specifications / Ramp Limits|MA15Specifications] |
+ |
+ More information on intranet page (Drawing): [LdmWiki:MA15] |