Removed lines 8-10 |
- [ma15b.jpg] |
- [ma15c.jpg] |
- [ma15d.jpg] |
At line 11 added 2 lines. |
+ [ma15d.jpg] |
+ |
Lines 21-22 were replaced by line 20 |
- At present, MA15 can only be used on RITA2. However, some |
- [Restrictions|MA15RitaRestrictions] have to be obeyed. |
+ !Restrictions |
Line 24 was replaced by lines 22-32 |
- [Omega Motor Change on Rita2] |
+ __after several quenches, operation above 12.5 without lambda stage is not recommended [see|http://samenv:8080/sample_environment/61]__ |
+ |
+ Due to force limits, restrictions are defined for the use of MA15. |
+ ||Instrument ||Allowed Max. Field [[Tesla] |
+ |RITA-2 |11.5 ... 14.9, see [MA15 Rita Restrictions] |
+ |TASP |14.9, see [MA15 TASP Restrictions] |
+ |EIGER |13, see [MA15 Eiger Restrictions] |
+ |ZEBRA |14.9 for the longer distance as defined for MA15 |
+ On all other instruments, MA15 can not yet be used. |
+ |
+ [Specifications / Ramp Limits|MA15Specifications] |
Line 26 was replaced by line 34 |
- [Specifications|MA15Specifications] |
+ More information on intranet page (Drawing): [LdmWiki:MA15] |
At line 27 added 2 lines. |
+ [Omega Motor Change on Rita2] |
+ |
Removed lines 45-51 |
- * The rotatable sample stick has a relatively bad thermal contact. Therefore the |
- minimal sample temperature is only about 2.5 K. It seems that increasing the |
- amount of exchange gas does not help until it begins to condensate at |
- the heat exchanger. If the scattering from the exchange gas does not |
- disturb the measurement, achieving something around 1.8 K should be |
- feasible. |
- |
Line 57 was replaced by lines 60-63 |
- |
+ |
+ [ma15b.jpg] |
+ [ma15a.jpg] |
+ [ma15c.jpg] |