Difference between version 43 and version 15:
At line 7 added 3 lines.
+
+ [ma15d.jpg]
+
Lines 17-18 were replaced by line 20
- At present, MA15 can only be used on RITA2. However, some
- [Restrictions|MA15RitaRestrictions] have to be obeyed.
+ !Restrictions
Line 20 was replaced by line 22
- [Specifications|MA15Specifications]
+ __after several quenches, operation above 12.5 without lambda stage is not recommended [see|http://samenv:8080/sample_environment/61]__
Line 22 was replaced by lines 24-34
- [Cooldown Procedure|MA15Cooldown]
+ Due to force limits, restrictions are defined for the use of MA15.
+ ||Instrument ||Allowed Max. Field [[Tesla]
+ |RITA-2 |11.5 ... 14.9, see [MA15 Rita Restrictions]
+ |TASP |14.9, see [MA15 TASP Restrictions]
+ |EIGER |13, see [MA15 Eiger Restrictions]
+ |ZEBRA |14.9 for the longer distance as defined for MA15
+ On all other instruments, MA15 can not yet be used.
+
+ [Specifications / Ramp Limits|MA15Specifications]
+
+ More information on intranet page (Drawing): [LdmWiki:MA15]
Line 24 was replaced by line 36
- !Solved Issues
+ [Omega Motor Change on Rita2]
Lines 26-28 were replaced by line 38
- * The automatic needle valve control did not work properly. It seems that this
- was a mechanical problem. The magnet was sent back to
- Oxford Instruments and is now back and operational.
+ [Cooldown Procedure|MA15Cooldown]
Lines 32-34 were replaced by lines 42-50
- * The rotatable sample stick has a relatively bad thermal contact. Therefore the
- minimal sample temperature is only about 2.5 K. It seems that putting more
- exchange gas does not elimate this problem.
+ * The automatic needle valve does not always work properly.
+ A mechanical problem was solved by Oxford Instruments.
+ It is still a problem that sometimes the needle valve bar
+ gets frozen. This problem might be avoided when keeping warm
+ the top of the needle valve during fill. However, at high
+ fields, the air dryer might not work!
+
+ * The nitrogen level gauge does not work. This has to be fixed, but
+ it seems that it may appear often.
At line 35 added 3 lines.
+ * During He-fill, LN2 has to be filled again in order to avoid
+ air sucking into the LN2 dewar.
+
Line 41 was replaced by lines 60-63
-
+
+ [ma15b.jpg]
+ [ma15a.jpg]
+ [ma15c.jpg]

Back to MA15, or to the Page History.