Diff: MA15

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
DilSampleMounting
Strayfields
DilutionCryostats
MagnetList
Spacerlist
Efield
EarlierProgress

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 43 and version 11:
Line 5 was replaced by line 5
- and the Danish National Science Foundation
+ and the Danish National Science Foundation (danish: [FNU|http://www.forsk.dk])
At line 7 added 3 lines.
+
+ [ma15d.jpg]
+
Lines 17-18 were replaced by lines 20-32
- At present, MA15 can only be used on RITA2. However, some
- [Restrictions|MA15RitaRestrictions] have to be obeyed.
+ !Restrictions
+
+ __after several quenches, operation above 12.5 without lambda stage is not recommended [see|http://samenv:8080/sample_environment/61]__
+
+ Due to force limits, restrictions are defined for the use of MA15.
+ ||Instrument ||Allowed Max. Field [[Tesla]
+ |RITA-2 |11.5 ... 14.9, see [MA15 Rita Restrictions]
+ |TASP |14.9, see [MA15 TASP Restrictions]
+ |EIGER |13, see [MA15 Eiger Restrictions]
+ |ZEBRA |14.9 for the longer distance as defined for MA15
+ On all other instruments, MA15 can not yet be used.
+
+ [Specifications / Ramp Limits|MA15Specifications]
Line 20 was replaced by line 34
- [Specifications|MA15Specifications]
+ More information on intranet page (Drawing): [LdmWiki:MA15]
At line 21 added 2 lines.
+ [Omega Motor Change on Rita2]
+
Lines 26-32 were replaced by lines 42-50
- * The automatic needle valve control does not work properly. It seems that this
- is a mechanical problem. The magnet was sent back to
- Oxford Instruments and should be back in the beginning of May 2005.
-
- * The rotatable sample stick has a relatively bad thermal contact. Therefore the
- minimal sample temperature is increased. It is not sure if putting more
- exchange gas can elimate this problem.
+ * The automatic needle valve does not always work properly.
+ A mechanical problem was solved by Oxford Instruments.
+ It is still a problem that sometimes the needle valve bar
+ gets frozen. This problem might be avoided when keeping warm
+ the top of the needle valve during fill. However, at high
+ fields, the air dryer might not work!
+
+ * The nitrogen level gauge does not work. This has to be fixed, but
+ it seems that it may appear often.
At line 33 added 3 lines.
+ * During He-fill, LN2 has to be filled again in order to avoid
+ air sucking into the LN2 dewar.
+
Line 39 was replaced by lines 60-63
-
+
+ [ma15b.jpg]
+ [ma15a.jpg]
+ [ma15c.jpg]

Back to MA15, or to the Page History.