Lines 8-38 were replaced by lines 8-14 |
- !Specifications |
- ||Magnet |
- |field at sample |14.9 Tesla |
- | " without lambda stage|13.5 Tesla |
- |operating current for 14.9 T|120 A |
- |maximum allowed vertical and horiontal magnetic forces|1000 N |
- |homogeneity over a 10 mm diameter sphere|0.5% |
- |field stability in persistent mode|0.01 % / hour |
- ||Variable Temperature Insert |
- |temperature range (no field dependend limitations)|1.6 ... 300 K |
- |temperature stability|+- 0.1 K |
- |automatic needle valve control only above|3 K |
- |exchange gas needed in sample space (measured at RT)|<15 mbar |
- ||Dimensions |
- |sample space diameter above magnet|49.5 mm |
- |diameter of sample space|20 mm |
- |max. useable sample height|20 mm |
- |split access angle|+- 2 deg |
- |sample height for +-2 deg access(visible by neutrons) |14 mm |
- |horizontal access angle|> 330 deg |
- |distance from base of cryostat\\to scattering plane|240 mm |
- |total radial aluminium window thickness|43.4 mm |
- |total radial aluminium in entrance window|8.6 mm |
- |entrance window size (radiussed corners 3 mm)|15 x 15 mm |
- |diameter for transfer tube|12 mm |
- ||Cryostat |
- |total weight including cryogens|650+-30 kg |
- |useable LHe capacity|55 l |
- |LHe boil off (static mode)|<350 cc/hr |
- |LN2 capacity|60 l |
- |typical LN2 boil off|600 cc/hr |
+ |
+ [Specifications|MA15Specifications] |
+ |
+ [Cooldown Procedure|MA15Cooldown] |
+ |
+ !Actual Problems |
+ |