Diff: MA15

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
DilSampleMounting
Strayfields
DilutionCryostats
MagnetList
Spacerlist
Efield
EarlierProgress

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 3 and version 2:
At line 15 added 5 lines.
+ ||Variable Temperature Insert
+ |temperature range (no field dependend limitations)|1.6 ... 300 K
+ |temperature stability|+- 0.1 K
+ |automatic needle valve control only above|3 K
+ |exchange gas needed in sample space (measured at RT)|<15 mbar
At line 16 added 1 line.
+ |sample space diameter above magnet|49.5 mm
Removed line 23
- |total weight including cryogens|650+-30 kg
At line 26 added 3 lines.
+ |diameter for transfer tube|12 mm
+ ||Cryostat
+ |total weight including cryogens|650+-30 kg
Removed lines 31-36
- |diameter for transfer tube|12 mm
- |temperature range (no field dependend limitations)|1.6 ... 300 K
- |temperature stability|+- 0.1 K
- |automatic needle valve control only above|3 K
- |exchange gas needed in sample space (measured at RT)|<15 mbar
- |sample space diameter above magnet|49.5 mm

Back to MA15, or to the Page History.