Line 26 was replaced by line 26 |
- !Solved Issues |
+ !Pending Issues |
Lines 28-30 were replaced by lines 28-33 |
- * The automatic needle valve control did not work properly. It seems that this |
- was a mechanical problem. The magnet was sent back to |
- Oxford Instruments and is now back and operational. |
+ * The automatic needle valve does not always work properly. |
+ A mechanical problem was solved by Oxford Instruments. |
+ It is still a problem that sometimes the needle valve bar |
+ gets frozen. This problem might be avoided when keeping warm |
+ the top of the needle valve during fill. However, at high |
+ fields, the air dryer might not work! |
Line 32 was replaced by lines 35-39 |
- !Pending Issues |
+ * The nitrogen level gauge does not work. This has to be fixed, but |
+ it seems that it may appear often. |
+ |
+ * During He-fill, LN2 has to be filled again in order to avoid |
+ air sucking into the LN2 dewar. |