Difference between version 17 and version 16:
Line 26 was replaced by line 26
- !Solved Issues
+ !Pending Issues
Lines 28-30 were replaced by lines 28-33
- * The automatic needle valve control did not work properly. It seems that this
- was a mechanical problem. The magnet was sent back to
- Oxford Instruments and is now back and operational.
+ * The automatic needle valve does not always work properly.
+ A mechanical problem was solved by Oxford Instruments.
+ It is still a problem that sometimes the needle valve bar
+ gets frozen. This problem might be avoided when keeping warm
+ the top of the needle valve during fill. However, at high
+ fields, the air dryer might not work!
Line 32 was replaced by lines 35-39
- !Pending Issues
+ * The nitrogen level gauge does not work. This has to be fixed, but
+ it seems that it may appear often.
+
+ * During He-fill, LN2 has to be filled again in order to avoid
+ air sucking into the LN2 dewar.

Back to MA15, or to the Page History.