| This is version 8.  
            It is not the current version, and thus it cannot be edited.[Back to current version]  
            [Restore this version]
 
 MA6 vertical field cryomagnet
 Restrictions
max. Forces on Magnet: 
 
| horizontal | 500N |  | vertical | 500N |  
 
| Instrument | Field Limit | Comment | 
|---|
 | DMC | 2.8 Tesla |  |  | MARS | 6 Tesla |  |  | TriCS | 6 Tesla |  |  | PREP | 6 Tesla | in its trolley, far enough from "Kabelkanal" |  | HRPT | 6 Tesla | (measured forces 140 N vertically and horizontally) |  
MA6 Specifications?
 
 Solved Issues
The following problems seem to be solved after repair at Oxford Instruments (Aug 2010)
 
 The magnet seems to have a cold leak, occuring only occasionaly at low temperatures.
  As a workaround the main vacuum has to be pumped constantly. (Jul 2009)
 The magnet sometimes quenches (only above 4.8 T)
 
 
 
      
 Attachments: 
        
        ma6_homogeneity.png