| Instrument | Field Limit | Comment | 
|---|---|---|
| DMC | 6 Tesla | |
| MARS | 6 Tesla | |
| TriCS | 6 Tesla | |
| PREP | 6 Tesla | in its trolley, far enough from "Kabelkanal" | 
| HRPT | 6 Tesla | (measured forces below 200 N vertically and horizontally, even for xy table) | 
| RITA II | 6 Tesla | with goniometer mounted (goniometer motors removed? - to be verified) | 
| EIGER | 6 Tesla | with goniometer mounted (including gonio motors) | 
| TASP | 6 Tesla | with goniometer mounted (including gonio motors) | 
On all other instruments, to be evaluated before first use.
max. Forces on Magnet:
| horizontal | 500N | 
| vertical | 500N | 
max. goniometer tilt: +-5 deg
More Information on intranet page (Drawing): LdmWiki:MA6 
| Field range | Training mode (slow) | Normal mode (fast) | 
| below 4.8 T | 0.1922 | 0.3844 | 
| 4.8 ... 6 T | 0.0641 | 0.3844 | 
| Guaranteed maximum central magnetic field | 6 Tesla | 
| Current for full field | 93.661 A | 
| Homogeneity over 15 mm diameter sphere | 1.0 % | 
| Magnet clear bore diameter | 50 mm | 
| Nominal inductance | 18.13 Henry | 
| Switch heater current for open state | 65 mA | 
| Current decay in persistent mode | < 1/10000 / h | 
| Distance of magnetic field centre for base of cryostat | 165 mm | 
| Neutron split | 40 mm | 
| Vertical neutron access on a 15 mm sphere | ±15 deg | 
| Horizontal neutron access | 340 deg | 
| RT | 77 K | 4.2 K | |
| Magnet coil | 48.4 | 39.4 | 0.1 | 
| Switch heater | 114.7 | 112.5 | 104.5 | 
| Spare switch heater | 112.7 | 110.4 | 103.5 | 
| Thermometer on coil (Allen Bradley) | 317.5 | 385 | 4560 | 
| Liquid He evaoration rate (needle vale close, no current) | <275 cm3/h | 
| Liquid N2 evaporation rate | 300 cm3/h | 
| Useful liquid He volume | 22 l | 
| Liquid N2 volume | 25 l | 
| Approximate weight | 150 kg | 
 
The following problems seem to be solved after repair at Oxford Instruments (Aug 2010)