MA6 vertical field cryomagnet

Restrictions

max. Forces on Magnet:

horizontal 500N
vertical 500N

Instrument Field Limit Comment
DMC 2.8 Tesla
MARS 6 Tesla
TriCS 6 Tesla
PREP 6 Tesla in its trolley, far enough from "Kabelkanal"
HRPT 6 Tesla (measured forces 140 N vertically and horizontally)

Specifications?

Solved Issues

The following problems seem to be solved after repair at Oxford Instruments (Aug 2010)

  • The magnet seems to have a cold leak, occuring only occasionaly at low temperatures. As a workaround the main vacuum has to be pumped constantly. (Jul 2009)
  • The magnet sometimes quenches (only above 4.8 T)