MA6 vertical field cryomagnet

Restrictions

Instrument Field Limit Comment
DMC 2.8 Tesla
MARS 6 Tesla
TriCS 6 Tesla
PREP 6 Tesla in its trolley, far enough from "Kabelkanal"
HRPT 6 Tesla (measured forces 140 N vertically and horizontally)

max. Forces on Magnet:

horizontal 500N
vertical 500N

Specifications

Energisation rates (Tesla/min)

Field rangeTraining mode (slow)Normal mode (fast)
below 4.8 T0.1922 0.3844
4.8 ... 6 T0.0641 0.3844

Main Coils

Guaranteed maximum central magnetic field6 Tesla
Current for full field 93.661 A
Homogeneity over 15 mm diameter sphere 1.0 %
Magnet clear bore diameter 50 mm
Nominal inductamce 18.13 Henry
Switch heater current for open state 65 mA
Current decay in persistent mode < 1/10000 / h
Distance of magnetic field centre for base of cryostat 165 mm
Neutron split 40 mm
Vertical neutron access on a 15 mm sphere±15 deg
Horizontal neutron access 340 deg

Resistance values (Ohm)

RT 77 K 4.2 K
Magnet coil 48.4 39.4 0.1
Switch heater 114.7112.5104.5
Spare switch heater 112.7110.4103.5
Thermometer on coil (Allen Bradley)317.5385 4560

General

Liquid He evaoration rate (needle vale close, no current)<275 cm3/h
Liquid N2 evaporation rate 300 cm3/h
Useful liquid He volume 22 l
Liquid N2 volume 25 l
Approximate weight 150 kg

Pending Issues

Some quenches appeared in 2011, we hope that can be solved by strictly obeying the training mode ramps.

Solved Issues

The following problems seem to be solved after repair at Oxford Instruments (Aug 2010)

  • The magnet seems to have a cold leak, occuring only occasionaly at low temperatures. As a workaround the main vacuum has to be pumped constantly. (Jul 2009)
  • The magnet sometimes quenches (only above 4.8 T)