This is version 10.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
MA6 vertical field cryomagnet
Restrictions
max. Forces on Magnet:
horizontal | 500N |
vertical | 500N |
Instrument | Field Limit | Comment |
DMC | 2.8 Tesla | |
MARS | 6 Tesla | |
TriCS | 6 Tesla | |
PREP | 6 Tesla | in its trolley, far enough from "Kabelkanal" |
HRPT | 6 Tesla | (measured forces 140 N vertically and horizontally) |
MA6 Specifications?
Pending Issues
Some quenches appeared in 2011, we hope that can be solved by strictly obeying the
training mode ramps.
Solved Issues
The following problems seem to be solved after repair at Oxford Instruments (Aug 2010)
- The magnet seems to have a cold leak, occuring only occasionaly at low temperatures.
As a workaround the main vacuum has to be pumped constantly. (Jul 2009)
- The magnet sometimes quenches (only above 4.8 T)
Attachments:
ma6_homogeneity.png
This particular version was published on 10-Jan-2012 12:17:17 UTC by MarkusZolliker.
|