MA6 vertical field cryomagnet

Restrictions

Instrument Field Limit Comment
DMC 2.5 Tesla
MARS 6 Tesla
TriCS 6 Tesla

Pending Issues

  • The magnet seems to have a cold leak, occuring only occasionaly at low temperatures. As a workaround the main vacuum has to be pumped constantly. (Jul 2009)

  • The magnet sometimes quenches (only above 4.8 T)



Attachments:  ma6_homogeneity.png 
This particular version was published on 25-Aug-2009 12:40:32 UTC by MarkusZolliker.