This is version 25.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
DIL Sample mounting
used in | | VARIOX | MA15 | MA11 | MA02 | MA09 | MA6 | MA7 | MA10 |
max. sample diameter* (incl. holder) [mm] | d | 40 | 14 | 14 | 21 | 40 | 21 | 32 | 40 |
sample space diameter [mm] | | 42.8 | 17 | 17 | 23.1 | 42.8 | 23.4 | 35.8 | 42.8 |
distance from DIL2 mixing chamber to beam center [mm] | l | 83+ | 296 | 296 | 342+-5 | 272 | 295+-5 | n.a. | ? |
distance from DIL3/DIL4 mixing chamber to beam center [mm] | l | 61+ | 296 | 296 | 342+-5 | 250 | 295+-5 | 350 | ? |
allowed size below beam center [mm] | b | | 32 | 32 | 60 | 24** | 65 | 59 | 32 |
(*) The maximum allowed size depends on the distance to the
next spacer, which is larger when using the weak link. Please contact
Markus Zolliker for designing sample holders slightly larger than the recommended max. sample diameter.
(**) Slightly higher values might be acceptable on request.
On the mixing chamber as well as at the end of the
sample holder/weak link there is an M6 hole, 6 mm deep. The (old) weak link
has a length of m = 245 mm.
For some experiments, it might be possible to fit 2 samples in the DIL insert, making possible to measure 2 samples in sequence
without removing the insert (saving one working day). For MA11 and MA15, the sample holder should look like the following picture.
Drawings of the DIL cans can be found at the intranet http://lns00/ldmwiki/Wiki.jsp?page=DIL
Attachments:
Dil2Samples.png
This particular version was published on 13-Sep-2018 08:47:54 UTC by MarkusZolliker.
|