| Lines 7-8 were replaced by lines 7-8 |
| - is a cone. Perpendicular to the field direction, the acces is +-6.7 deg on |
| - a 15 mm sample. |
| + is a cone. Perpendicular to the field direction, the access is +-6.7 deg on |
| + a 15 mm sample, or 65 mm parallel access. |
| At line 27 added 1 line. |
| + [MA7/asysym.jpg] |
| Line 30 was replaced by line 31 |
| - For changing for symmetric mode to asymmetric and vice versa: |
| + For changing from symmetric mode to asymmetric mode and vice versa: |
| At line 32 added 3 lines. |
| + ** make sure that the magnetic field is zero (IPS: shown current is 0 AND "persistent mode" LED is off) |
| + ** switch the magnet power supply IPS off |
| + ** open the cap using the screws marked on the picture above |
| At line 34 added 2 lines. |
| + ** close the cap |
| + ** switch the magnet power supply on |
| Line 49 was replaced by line 55 |
| - |S11 (coil inductance) | 112 | 78.7 |
| + |S11 (coil inductance) | 112 | 78.8 |