Line 1 was replaced by line 1 |
- !!MA7 |
+ !!6.8 Tesla Horizontal Field Cryomagnet MA7 |
Line 3 was replaced by lines 3-15 |
- This new 6.8 Tesla horizontal field cryomagnet has strong stray fields due |
+ The horizontal high field cryomagnet has a restricted neutron access. |
+ There are 4 windows of 45 deg each, valid for a sample of 15 mm diameter. |
+ For larger samples, the opening angle for full access is sligthly smaller. |
+ In the field direction, the vertical access is +- 22.5, in fact, the access |
+ is a cone. Perpendicular to the field direction, the access is +-6.7 deg on |
+ a 15 mm sample, or 65 mm parallel access. |
+ |
+ [MA7/MA7coils.png] |
+ |
+ For access with polarized neutrons, an asymmetric mode (max. 6.2 Tesla) |
+ is available. This is needed when a beam perpendicular to the field is used. |
+ |
+ This cryomagnet has strong stray fields due |
Line 8 was replaced by lines 20-24 |
- !Restrictions |
+ [MA7 Specifications] |
+ |
+ More information on intranet page (drawing): [LdmWiki:MA7] |
+ |
+ [MA7 Restrictions] |
Lines 10-11 were replaced by line 26 |
- Provisional force measuremnents on RITA-II have shown that |
- the operation of MA7 on Rita-II is allowed: |
+ [MA7/MA7pict.jpg] |
Lines 13-23 were replaced by lines 28-58 |
- * between 3.0 and 18.0 meV (a2 >= 37 deg) at a distance of 450 mm or more, up to 5 Tesla |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 450 mm or more, up to 5.5 Tesla |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 540 mm or more, up to 6.8 Tesla |
- |
- For any other configuration force tests have to be done. |
- __Anyway, these measurments should be confirmed when an |
- approriate adapter plate is avaliable.__ |
- |
- [MA15RitaRestrictions/SampleTableDistance.jpg]\\ |
- How the distance has to be measured. |
- In this example the distance is around 400 mm. |
+ [MA7/asysym.jpg] |
+ !Asymetric mode |
+ |
+ For changing from symmetric mode to asymmetric mode and vice versa: |
+ |
+ * on the top of the magnet under the aluminium cyclinder: |
+ ** make sure that the magnetic field is zero (IPS: shown current is 0 AND "persistent mode" LED is off) |
+ ** switch the magnet power supply IPS off |
+ ** open the cap using the screws marked on the picture above |
+ ** switch the cable from the SYM pin to ASY pin (or vice versa) |
+ ** the cable connected to the COM pin must always stay connected |
+ ** close the cap |
+ ** switch the magnet power supply on |
+ * change the configuration of the IPS: |
+ ** press RAISE+LOWER+LOC/REM simultanously ("tESt" and "t 00" appears) |
+ ** press LOWER to go to "t 07", and then LOC/REM (""SUP" and "S 00" appears) |
+ ** press LOWER 5 times to go to "S 11" and then press LOC/REM (the induction value appears) |
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done |
+ ** press RAISE 7 times to go to "S 02" and then press LOC/REM (the value for max. current appears) |
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done |
+ ** press LOWER once to go to "S 01" and then press LOC/REM (the value for max. current appears) |
+ ** press LOC/REM again (the value for max. field appears) |
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done |
+ ** press LOWER once to go to "S 00" and then press LOC/REM ("tESt" and "t 00" appears) |
+ ** press LOC/REM again ("Stor.y" appears) and again |
+ |
+ ||Parameter || Value for Sym || Value for Asym |
+ |S11 (coil inductance) | 112 | 78.7 |
+ |S02 (max. current) | 96.636 | 100.657 |
+ |S01 (max. field: second value)| 6.8 | 6.2 |
+ |