Lines 7-8 were replaced by lines 7-8 |
- is a cone. Perpendicular to the field direction, the acces is +-6.7 deg on |
- a 15 mm sample. |
+ is a cone. Perpendicular to the field direction, the access is +-6.7 deg on |
+ a 15 mm sample, or 65 mm parallel access. |
Line 31 was replaced by line 31 |
- For changing for symmetric mode to asymmetric and vice versa: |
+ For changing from symmetric mode to asymmetric mode and vice versa: |
At line 33 added 3 lines. |
+ ** make sure that the magnetic field is zero (IPS: shown current is 0 AND "persistent mode" LED is off) |
+ ** switch the magnet power supply IPS off |
+ ** open the cap using the screws marked on the picture above |
At line 35 added 2 lines. |
+ ** close the cap |
+ ** switch the magnet power supply on |