+ In the field direction, the vertical access is +- 22.5, in fact, the access
+ is a cone. Perpendicular to the field direction, the access is +-6.7 deg on
+ a 15 mm sample, or 65 mm parallel access.
At line 16 added 4 lines.
+ [MA7 Specifications]
+
+ More information on intranet page (drawing): [LdmWiki:MA7]
+
At line 20 added 1 line.
+ [MA7/asysym.jpg]
Line 23 was replaced by line 31
- For changing for symmetric mode to asymmetric and vice versa:
+ For changing from symmetric mode to asymmetric mode and vice versa:
Line 25 was replaced by lines 33-40
- * change a cable on the top of the magnet under the aluminium cyclinder
+ * on the top of the magnet under the aluminium cyclinder:
+ ** make sure that the magnetic field is zero (IPS: shown current is 0 AND "persistent mode" LED is off)
+ ** switch the magnet power supply IPS off
+ ** open the cap using the screws marked on the picture above
+ ** switch the cable from the SYM pin to ASY pin (or vice versa)
+ ** the cable connected to the COM pin must always stay connected
+ ** close the cap
+ ** switch the magnet power supply on

Back to MA7, or to the Page History.