At line 4 added 4 lines.
+ For larger samples, the opening angle for full access is sligthly smaller.
+ In the field direction, the vertical access is +- 22.5, in fact, the access
+ is a cone. Perpendicular to the field direction, the access is +-6.7 deg on
+ a 15 mm sample, or 65 mm parallel access.
At line 15 added 4 lines.
+ [MA7 Specifications]
+
+ More information on intranet page (drawing): [LdmWiki:MA7]
+
At line 19 added 1 line.
+ [MA7/asysym.jpg]
Line 22 was replaced by line 31
- For changing for symmetric mode to asymmetric and vice versa:
+ For changing from symmetric mode to asymmetric mode and vice versa:
Lines 24-25 were replaced by lines 33-52
- * change a cable on the top of the magnet under the aluminium cyclinder
- * change the configuration of the IPS, (Test 07):
+ * on the top of the magnet under the aluminium cyclinder:
+ ** make sure that the magnetic field is zero (IPS: shown current is 0 AND "persistent mode" LED is off)
+ ** switch the magnet power supply IPS off
+ ** open the cap using the screws marked on the picture above
+ ** switch the cable from the SYM pin to ASY pin (or vice versa)
+ ** the cable connected to the COM pin must always stay connected
+ ** close the cap
+ ** switch the magnet power supply on
+ * change the configuration of the IPS:
+ ** press RAISE+LOWER+LOC/REM simultanously ("tESt" and "t 00" appears)
+ ** press LOWER to go to "t 07", and then LOC/REM (""SUP" and "S 00" appears)
+ ** press LOWER 5 times to go to "S 11" and then press LOC/REM (the induction value appears)
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done
+ ** press RAISE 7 times to go to "S 02" and then press LOC/REM (the value for max. current appears)
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done
+ ** press LOWER once to go to "S 01" and then press LOC/REM (the value for max. current appears)
+ ** press LOC/REM again (the value for max. field appears)
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done
+ ** press LOWER once to go to "S 00" and then press LOC/REM ("tESt" and "t 00" appears)
+ ** press LOC/REM again ("Stor.y" appears) and again
At line 27 added 1 line.
+ |S11 (coil inductance) | 112 | 78.7
Removed line 30
- |S11 (coil inductance) | 112 | 78.7

Back to MA7, or to the Page History.