Removed lines 3-4 |
- [MA7/MA7pict.jpg] |
- |
At line 6 added 4 lines. |
+ For larger samples, the opening angle for full access is sligthly smaller. |
+ In the field direction, the vertical access is +- 22.5, in fact, the access |
+ is a cone. Perpendicular to the field direction, the access is +-6.7 deg on |
+ a 15 mm sample, or 65 mm parallel access. |
Line 18 was replaced by lines 20-29 |
- !Restrictions on RITA-II |
+ [MA7 Specifications] |
+ |
+ More information on intranet page (drawing): [LdmWiki:MA7] |
+ |
+ [MA7 Restrictions] |
+ |
+ [MA7/MA7pict.jpg] |
+ |
+ [MA7/asysym.jpg] |
+ !Asymetric mode |
Lines 20-21 were replaced by line 31 |
- Provisional force measuremnents on RITA-II have shown that |
- the operation of MA7 on Rita-II is allowed: |
+ For changing from symmetric mode to asymmetric mode and vice versa: |
Lines 23-85 were replaced by lines 33-58 |
- * between 3.0 and 18.0 meV (a2 >= 37 deg) at a distance of 450 mm or more, up to 4 Tesla |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 450 mm or more, up to 4.4 Tesla |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 540 mm or more, up to 5.5 Tesla |
- |
- For any other configuration force tests have to be done. |
- As soon as the approriate adapter plate is avaliable, we should |
- repeat the force measurements. |
- |
- Original data: |
- ||a2 [[deg] || distance [[mm] ||field [[Tesla] || force [[N] || extrapolated field [[Tesla] for 500 N || allowed field with security margin |
- | 56 | 540 | 4.8 | 230 | 6.8 | 5.5 |
- | 56 | 450 | 3.4 | 180 | 5.5 | 4.4 |
- | 37 | 450 | 3.4 | 220 | 5 | 4 |
- |
- [MA15RitaRestrictions/SampleTableDistance.jpg]\\ |
- How the distance has to be measured. |
- In this example the distance is around 400 mm. |
- |
- !Restrictions on AMOR |
- |
- Provisional force measuremnents on AMOR have shown that |
- the operation of MA7 on AMOR is allowed with a field perpendicular |
- to the beam at least up to 4 Tesla. |
- The distance to the shielding wall between AMOR and SANS-II has |
- to be at least 540 mm, measured from the aluminium window |
- (or about 870 mm to the magnet centre). |
- |
- ''Torque Calculations'' |
- |
- In general, the forces are not only radial, but also tangential, |
- inducing a torque onto the magnet. For estimating the forces |
- on the fixation, we need to take into account the distance of |
- the fixation points from the coil center. |
- |
- For MA7, with a mean coil radius of about 100 m and fixation points |
- at about 150 mm for the coil center, we get the following: |
- |
- Ro: object distance (mm)\\ |
- phi: angle (deg) between field direction and direction of object, for maximum force on fixation\\ |
- Fp: force when object on magnet axis (in field direction) |
- Ft: maximum total force on fixation |
- |
- ||Ro ||phi ||Ft / Fp |
- |<600 | 0 | 1.0 |
- | 800 | 20 | 1.04 |
- | 1000 | 27 | 1.13 |
- | 1200 | 32 | 1.25 |
- |
- For the distance to the shielding wall we have to add 1/5 of the edge |
- distance for the mean distance of the material that has to be taken |
- into account (870 * 6/5 = 1044 mm). |
- |
- Measured force at 3.4 Tesla: 180 N.\\ |
- Extrapolated to 5 Tesla: 390 N.\\ |
- Factor 1.2 for including torque: 470 N.\\ |
- With a reasonable security margin, we allow to go up to 4 Tesla. |
- |
- !Restrictions on TASP |
- |
- ||A2 || distance sample - \\monochromator shielding surface|| Allowed field |
- |-90 < A2 < -41 (Ei=14.7 meV)|>=122 cm | 3.3 Tesla |
- |-90 < A2 < -56 (Ei=8 meV)|>=108 cm | 4 Tesla |
- |-90 < A2 < -82 (Ei=4.1 meV)|>=108 cm | 5 Tesla |
+ * on the top of the magnet under the aluminium cyclinder: |
+ ** make sure that the magnetic field is zero (IPS: shown current is 0 AND "persistent mode" LED is off) |
+ ** switch the magnet power supply IPS off |
+ ** open the cap using the screws marked on the picture above |
+ ** switch the cable from the SYM pin to ASY pin (or vice versa) |
+ ** the cable connected to the COM pin must always stay connected |
+ ** close the cap |
+ ** switch the magnet power supply on |
+ * change the configuration of the IPS: |
+ ** press RAISE+LOWER+LOC/REM simultanously ("tESt" and "t 00" appears) |
+ ** press LOWER to go to "t 07", and then LOC/REM (""SUP" and "S 00" appears) |
+ ** press LOWER 5 times to go to "S 11" and then press LOC/REM (the induction value appears) |
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done |
+ ** press RAISE 7 times to go to "S 02" and then press LOC/REM (the value for max. current appears) |
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done |
+ ** press LOWER once to go to "S 01" and then press LOC/REM (the value for max. current appears) |
+ ** press LOC/REM again (the value for max. field appears) |
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done |
+ ** press LOWER once to go to "S 00" and then press LOC/REM ("tESt" and "t 00" appears) |
+ ** press LOC/REM again ("Stor.y" appears) and again |
+ |
+ ||Parameter || Value for Sym || Value for Asym |
+ |S11 (coil inductance) | 112 | 78.7 |
+ |S02 (max. current) | 96.636 | 100.657 |
+ |S01 (max. field: second value)| 6.8 | 6.2 |
+ |