Difference between version 44 and version 43:
Line 25 was replaced by lines 25-36
- * change the configuration of the IPS, (Test 07):
+ * change the configuration of the IPS:
+ ** press RAISE+LOWER+LOC/REM simultanously ("tESt" and "t 00" appears)
+ ** press LOWER to go to "t 07", and then LOC/REM (""SUP" and "S 00" appears)
+ ** press LOWER 5 times to go to "S 11" and then press LOC/REM (the induction value appears)
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done
+ ** press RAISE 7 times to go to "S 02" and then press LOC/REM (the value for max. current appears)
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done
+ ** press LOWER once to go to "S 01" and then press LOC/REM (the value for max. current appears)
+ ** press LOC/REM again (the value for max. field appears)
+ ** press and hold LOC/REM and change the value with RAISE or LOWER, release LOC/REM when done
+ ** press LOWER once to go to "S 00" and then press LOC/REM ("tESt" and "t 00" appears)
+ ** press LOC/REM again ("Stor.y" appears) and again

Back to MA7, or to the Page History.