Line 20 was replaced by line 20 |
- !Restrictions on RITA-II |
+ !Asymetric mode |
Lines 22-23 were replaced by line 22 |
- Provisional force measuremnents on RITA-II have shown that |
- the operation of MA7 on Rita-II is allowed: |
+ For changing for symmetric mode to asymmetric and vice versa: |
Lines 25-91 were replaced by lines 24-31 |
- * between 3.0 and 18.0 meV (a2 >= 37 deg) at a distance of 450 mm or more, __up to 4 Tesla__ |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 450 mm or more, __up to 4.4 Tesla__ |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 540 mm or more, __up to 5.5 Tesla__ (we had a quench at 6 Tesla!) |
- For any other configuration force tests have to be done. |
- As soon as the appropriate adapter plate is avaliable, we should |
- repeat the force measurements. |
- |
- [MA7 Original Data] |
- |
- [MA15RitaRestrictions/SampleTableDistance.jpg]\\ |
- How the distance has to be measured. |
- In this example the distance is around 400 mm. |
- |
- !Restrictions on AMOR |
- |
- Provisional force measuremnents on AMOR have shown that |
- the operation of MA7 on AMOR is allowed with a field perpendicular |
- to the beam at least up to 4 Tesla. |
- The distance to the shielding wall between AMOR and SANS-II has |
- to be at least 540 mm, measured from the aluminium window |
- (or about 870 mm to the magnet centre). |
- |
- ''Torque Calculations'' |
- |
- In general, the forces are not only radial, but also tangential, |
- inducing a torque onto the magnet. For estimating the forces |
- on the fixation, we need to take into account the distance of |
- the fixation points from the coil center. |
- |
- For MA7, with a mean coil radius of about 100 m and fixation points |
- at about 150 mm for the coil center, we get the following: |
- |
- Ro: object distance (mm)\\ |
- phi: angle (deg) between field direction and direction of object, for maximum force on fixation\\ |
- Fp: force when object on magnet axis (in field direction) |
- Ft: maximum total force on fixation |
- |
- ||Ro ||phi ||Ft / Fp |
- |<600 | 0 | 1.0 |
- | 800 | 20 | 1.04 |
- | 1000 | 27 | 1.13 |
- | 1200 | 32 | 1.25 |
- |
- For the distance to the shielding wall we have to add 1/5 of the edge |
- distance for the mean distance of the material that has to be taken |
- into account (870 * 6/5 = 1044 mm). |
- |
- Measured force at 3.4 Tesla: 180 N.\\ |
- Extrapolated to 5 Tesla: 390 N.\\ |
- Factor 1.2 for including torque: 470 N.\\ |
- With a reasonable security margin, we allow to go up to 4 Tesla. |
- |
- !Restrictions on TASP |
- |
- ||A2 || distance sample - \\monochromator shielding surface|| Allowed field |
- |-90 < A2 < -41 (Ei=14.7 meV)|>=122 cm | 3.3 Tesla |
- |-90 < A2 < -56 (Ei=8 meV)|>=108 cm | 4 Tesla |
- |-90 < A2 < -82 (Ei=4.1 meV)|>=108 cm | 5 Tesla |
- |
- !Restrictions on SANS |
- |
- Allowed up to maximum field independend of the field direction. |
- |
- Force measurements: |
- ||field direction || field || horizontal force || vertical force |
- |parallel to beam | 4.9 Tesla | 80 N | 340 N |
- |perpendicular to beam | 4.9 Tesla | 110 N | 180 N |
+ * change a cable on the top of the magnet under the aluminium cyclinder |
+ * change the configuration of the IPS, (Test 07): |
+ |
+ ||Parameter || Value for Sym || Value for Asym |
+ |S02 (max. current) | 96.636 | 100.657 |
+ |S01 (max. field: second value)| 6.8 | 6.2 |
+ |S11 (coil inductance) | 112 | 78.7 |
+ |