Lines 13-19 were replaced by lines 13-19 |
- * between 3.0 and 18.0 meV (a2 >= 37 deg) at a distance of 450 mm or more, up to 5 Tesla |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 450 mm or more, up to 5.5 Tesla |
- * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 540 mm or more, up to 6.8 Tesla |
- |
- For any other configuration force tests have to be done. |
- __Anyway, these measurments should be confirmed when an |
- approriate adapter plate is avaliable.__ |
+ * between 3.0 and 18.0 meV (a2 >= 37 deg) at a distance of 450 mm or more, up to 4 Tesla |
+ * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 450 mm or more, up to 4.4 Tesla |
+ * between 3.0 and 8.0 meV (a2 >= 56 deg) at a distance of 540 mm or more, up to 5.5 Tesla |
+ |
+ For any other configuration force tests have to be done. |
+ As soon as the approriate adapter plate is avaliable, we should |
+ repeat the force measurments. |
Lines 22-25 were replaced by lines 22-25 |
- ||a2 [[deg] || distance [[mm] ||field [[Tesla] || force [[N] |
- | 56 | 540 | 4.8 | 230 |
- | 56 | 450 | 3.4 | 180 |
- | 37 | 450 | 3.4 | 220 |
+ ||a2 [[deg] || distance [[mm] ||field [[Tesla] || force [[N] || extrapolated field [[Tesla] for 500 N || allowed field with security margin |
+ | 56 | 540 | 4.8 | 230 | 6.8 | 5.5 |
+ | 56 | 450 | 3.4 | 180 | 5.5 | 4.4 |
+ | 37 | 450 | 3.4 | 220 | 5 | 4 |