| Line 9 was replaced by line 9 | 
| - |DMC          |2.8 Tesla     | | 
| + |DMC          |6 Tesla       | | 
| Line 14 was replaced by lines 14-17 | 
| - |RITA II      |6 Tesla       | with goniometer mounted!!!! | 
| + |RITA II      |6 Tesla       | with goniometer mounted (goniometer motors removed? - to be verified) | 
| + |EIGER        |6 Tesla       | with goniometer mounted (including gonio motors) | 
| + |TASP         |6 Tesla       | with goniometer mounted (including gonio motors) | 
| + On EIGER the vertical force was less than 80 N at 4.25 Tesla -> no problem. (force test 2020-09-14 by MZ) | 
| At line 15 added 2 lines. | 
| + On all other instruments, to be evaluated before first use. | 
| + | 
| At line 19 added 3 lines. | 
| + max. goniometer tilt: | 
| + +-5 deg | 
| + | 
| At line 21 added 2 lines. | 
| + More Information on intranet page (Drawing): [LdmWiki:MA6] | 
| + | 
| Line 34 was replaced by line 44 | 
| - |Nominal inductamce                       |18.13 Henry | 
| + |Nominal inductance                       |18.13 Henry | 
| Line 58 was replaced by lines 68-69 | 
| - !Pending Issues | 
| + !Field Homogeneity | 
| + [MA6/ma6_homogeneity.png] | 
| Removed lines 60-62 | 
| - Some quenches appeared in 2011, we hope that can be solved by strictly obeying the | 
| - training mode ramps. | 
| - | 
| Line 68 was replaced by lines 76-78 | 
| - * The magnet sometimes quenches (only above 4.8 T) | 
| + * The magnet sometimes quenches (only above 4.8 T) -> solved by strictly obeying the | 
| +   training mode ramps. | 
| + |