| At line 2 added 3 lines. | 
| + !!Sample space dimensions | 
| + see [Magnet List] | 
| + | 
| Line 6 was replaced by line 9 | 
| - |DMC          |2.8 Tesla     | | 
| + |DMC          |6 Tesla       | | 
| Line 10 was replaced by lines 13-19 | 
| - |HRPT         |6 Tesla       | (measured forces 140 N vertically and horizontally) | 
| + |HRPT         |6 Tesla       | (measured forces below 200 N vertically and horizontally, even for xy table) | 
| + |RITA II      |6 Tesla       | with goniometer mounted (goniometer motors removed? - to be verified) | 
| + |EIGER        |6 Tesla       | with goniometer mounted (including gonio motors) | 
| + |TASP         |6 Tesla       | with goniometer mounted (including gonio motors) | 
| + On EIGER the vertical force was less than 80 N at 4.25 Tesla -> no problem. (force test 2020-09-14 by MZ) | 
| + | 
| + On all other instruments, to be evaluated before first use. | 
| At line 15 added 3 lines. | 
| + max. goniometer tilt: | 
| + +-5 deg | 
| + | 
| At line 17 added 2 lines. | 
| + More Information on intranet page (Drawing): [LdmWiki:MA6] | 
| + | 
| Line 30 was replaced by line 44 | 
| - |Nominal inductamce                       |18.13 Henry | 
| + |Nominal inductance                       |18.13 Henry | 
| Lines 54-57 were replaced by lines 68-69 | 
| - !Pending Issues | 
| - | 
| - Some quenches appeared in 2011, we hope that can be solved by strictly obeying the | 
| - training mode ramps. | 
| + !Field Homogeneity | 
| + [MA6/ma6_homogeneity.png] | 
| Line 64 was replaced by lines 76-78 | 
| - * The magnet sometimes quenches (only above 4.8 T) | 
| + * The magnet sometimes quenches (only above 4.8 T) -> solved by strictly obeying the | 
| +   training mode ramps. | 
| + |