| Line 1 was replaced by line 1 | 
| - !!MA6 vertical field cryomagnet | 
| + !!!MA6 vertical field cryomagnet | 
| Line 3 was replaced by lines 3-4 | 
| - !Restrictions | 
| + !!Sample space dimensions | 
| + see [Magnet List] | 
| Lines 5-7 were replaced by line 6 | 
| - max. Forces on Magnet: | 
| - |horizontal |500N | 
| - |vertical |500N | 
| + !!Restrictions | 
| Line 10 was replaced by line 9 | 
| - |DMC          |2.8 Tesla     | | 
| + |DMC          |6 Tesla       | | 
| Line 14 was replaced by lines 13-26 | 
| - |HRPT         |6 Tesla       | (measured forces 140 N vertically and horizontally) | 
| + |HRPT         |6 Tesla       | (measured forces below 200 N vertically and horizontally, even for xy table) | 
| + |RITA II      |6 Tesla       | with goniometer mounted (goniometer motors removed? - to be verified) | 
| + |EIGER        |6 Tesla       | with goniometer mounted (including gonio motors) | 
| + |TASP         |6 Tesla       | with goniometer mounted (including gonio motors) | 
| + On EIGER the vertical force was less than 80 N at 4.25 Tesla -> no problem. (force test 2020-09-14 by MZ) | 
| + | 
| + On all other instruments, to be evaluated before first use. | 
| + | 
| + max. Forces on Magnet: | 
| + |horizontal |500N | 
| + |vertical |500N | 
| + | 
| + max. goniometer tilt: | 
| + +-5 deg | 
| Line 16 was replaced by line 28 | 
| - [MA6 Specifications] | 
| + !!Specifications | 
| Line 18 was replaced by line 30 | 
| - !Pending Issues | 
| + More Information on intranet page (Drawing): [LdmWiki:MA6] | 
| Lines 20-21 were replaced by line 32 | 
| - Some quenches appeared in 2011, we hope that can be solved by strictly obeying the | 
| - training mode ramps. | 
| + !Energisation rates (Tesla/min) | 
| At line 22 added 37 lines. | 
| + |Field range|Training mode (slow)|Normal mode (fast) | 
| + |below 4.8 T|0.1922  |0.3844 | 
| + |4.8 ... 6 T|0.0641  |0.3844 | 
| + | 
| + !Main Coils | 
| + | 
| + |Guaranteed maximum central magnetic field|6 Tesla | 
| + |Current for full field                   |93.661 A | 
| + |Homogeneity over 15 mm diameter sphere   |1.0 % | 
| + |Magnet clear bore diameter               |50 mm | 
| + |Nominal inductance                       |18.13 Henry | 
| + |Switch heater current for open state     |65 mA | 
| + |Current decay in persistent mode         |< 1/10000 / h | 
| + |Distance of magnetic field centre for base of cryostat| 165 mm | 
| + |Neutron split                            |40 mm | 
| + |Vertical neutron access on a 15 mm sphere|±15 deg | 
| + |Horizontal neutron access                |340 deg | 
| + | 
| + !Resistance values (Ohm) | 
| + | 
| + |                                   |RT   |77 K |4.2 K | 
| + |Magnet coil                        |48.4 |39.4 |0.1 | 
| + |Switch heater                      |114.7|112.5|104.5 | 
| + |Spare switch heater                |112.7|110.4|103.5 | 
| + |Thermometer on coil (Allen Bradley)|317.5|385  |4560 | 
| + | 
| + !General | 
| + | 
| + |Liquid He evaoration rate (needle vale close, no current)|<275 cm3/h | 
| + |Liquid N2 evaporation rate |300 cm3/h | 
| + |Useful liquid He volume | 22 l | 
| + |Liquid N2 volume | 25 l | 
| + |Approximate weight | 150 kg | 
| + | 
| + !Field Homogeneity | 
| + [MA6/ma6_homogeneity.png] | 
| + | 
| Line 28 was replaced by lines 76-78 | 
| - * The magnet sometimes quenches (only above 4.8 T) | 
| + * The magnet sometimes quenches (only above 4.8 T) -> solved by strictly obeying the | 
| +   training mode ramps. | 
| + |