| Removed lines 5-8 | 
| - max. Forces on Magnet: | 
| - |horizontal |500N | 
| - |vertical |500N | 
| - | 
| Line 16 was replaced by lines 12-52 | 
| - [MA6 Specifications] | 
| + max. Forces on Magnet: | 
| + |horizontal |500N | 
| + |vertical |500N | 
| + | 
| + !!Specifications | 
| + | 
| + !Energisation rates (Tesla/min) | 
| + | 
| + |Field range|Training mode (slow)|Normal mode (fast) | 
| + |below 4.8 T|0.1922  |0.3844 | 
| + |4.8 ... 6 T|0.0641  |0.3844 | 
| + | 
| + !Main Coils | 
| + | 
| + |Guaranteed maximum central magnetic field|6 Tesla | 
| + |Current for full field                   |93.661 A | 
| + |Homogeneity over 15 mm diameter sphere   |1.0 % | 
| + |Magnet clear bore diameter               |50 mm | 
| + |Nominal inductamce                       |18.13 Henry | 
| + |Switch heater current for open state     |65 mA | 
| + |Current decay in persistent mode         |< 1/10000 / h | 
| + |Distance of magnetic field centre for base of cryostat| 165 mm | 
| + |Neutron split                            |40 mm | 
| + |Vertical neutron access on a 15 mm sphere|±15 deg | 
| + |Horizontal neutron access                |340 deg | 
| + | 
| + !Resistance values (Ohm) | 
| + | 
| + |                                   |RT   |77 K |4.2 K | 
| + |Magnet coil                        |48.4 |39.4 |0.1 | 
| + |Switch heater                      |114.7|112.5|104.5 | 
| + |Spare switch heater                |112.7|110.4|103.5 | 
| + |Thermometer on coil (Allen Bradley)|317.5|385  |4560 | 
| + | 
| + !General | 
| + | 
| + |Liquid He evaoration rate (needle vale close, no current)|<275 cm3/h | 
| + |Liquid N2 evaporation rate |300 cm3/h | 
| + |Useful liquid He volume | 22 l | 
| + |Liquid N2 volume | 25 l | 
| + |Approximate weight | 150 kg |