Difference between version 36 and version 2:
Lines 1-6 were replaced by lines 1-11
- !DIL1/DIL2/DIL3 Sample mounting
- |used in| ||MA15||MA11||MA02||MA09
- |sample space diameter [[mm]| |17|17|23.6|45
- |max. sample diameter [[mm]|d|14*|14*|22|40
- |distance from M6 thread to beam center [[mm]\\adjustable|a|51+-10|51+-10|<=97|<=194
- |allowed size below beam center [[mm]|b|32|32|60|24**
+ !DIL Sample mounting
+
+ See also [Calculator for Sample Mounting in DIL|Print:CalculatorForSampleMountingInDil]
+
+ |used in | ||[VARIOX]\\out of service||[ORI7]||[MA15] ||[MA11] ||[MA02]||[MA6]||[MA7]||[MA10]|[MB11]|[VarioxB]||[MA09]\\offline use only
+ |max. sample diameter* (incl. holder) [[mm] |d|40 |40 |14 |14 |21 |21 |32 |40 |77 |77 |40
+ |sample space diameter [[mm] | |42.8|42.8 |17 |17 |23.1 |23.4 |35.8 |42.8 |85.8 (77.4) |85.8 (77.4) |42.8
+ |distance from [DIL2] mixing chamber to beam center [[mm] |l|83+|83+ |296 |296 |342+-5 |295+-5 |n.a. |? | | |272
+ |distance from [DIL3]/[DIL4] mixing chamber to beam center [[mm]|l|61+|61+ |296 |296 |342+-5 |295+-5 |350 |? | | |250
+ |distance from [DIL5] mixing chamber to beam center [[mm] |l| | | | | | | | | |210+-10 |210+-10
+ |allowed size below beam center [[mm]|b| | |32 |32 |60 |65 |59 |32 |150 |150 |24**
Lines 8-9 were replaced by lines 13-14
- next spacer, which is larger when using the weak link. Please contact
- [Markus Zolliker] for designing sample holders larger than 14 mm in diameter.
+ next [spacer|Spacerlist], which is larger when using the weak link. Please contact
+ [Markus Zolliker] for designing sample holders slightly larger than the recommended max. sample diameter.
Lines 13-14 were replaced by lines 18-27
- sample holder there is an M6 hole. The sample holder
- has normally a length of m = 245 mm.
+ sample holder/weak link there is an M6 hole, 6 mm deep. The weak link
+ has a length of 30 mm.
+
+ For some experiments, it might be possible to fit 2 samples in the DIL insert, making possible to measure 2 samples in sequence
+ without removing the insert (saving one working day). For MA11 and MA15, the sample holder should look like the following picture.
+
+ [DilSampleMounting/Dil2Samples.png]
+
+
+ Drawings of the DIL cans can be found at the intranet [http://lns00/ldmwiki/Wiki.jsp?page=DIL]

Back to DilSampleMounting, or to the Page History.