Preparing the Kelvinox Dilution and Gas Handling System

  • pump the traps (heat out each for 30 min, but pump longer)
  • pump out the condenser and pumping line, on the gas handling system and on the stick (all volumes might be connected)

After some hours (or one night)

  • make the seal with clean fingers (a thick layer on the stick, only a small amount inside the aluminium tube)
  • put the tube, turning it a little on the seal and start pumping with the leak detector.
  • leak check
  • put into LN2 for precooling, put a small amount of He (1 mbar or less) into the IVC.
  • if there is no hurry nad the sample must be cooled slowly, the stick might be put directly into the cryostat (without LN2 precooling).

...



This particular version was published on 11-Aug-2005 14:00:13 UTC by MarkusZolliker.