DIL Sample mounting

used in VARIOXMA15 MA11 MA02MA09MA6 MA7
sample space diameter [mm] 4517 17 23.6 45 23.6 36
max. sample diameter [mm]d4014* 14* 22 40 22 32
distance from DIL1 mixing chamber to beam center [mm]l83n.a. n.a. 339 272 n.a. n.a.
distance from DIL2 mixing chamber to beam center [mm]l83296 296 339 272 294 n.a.
distance from DIL3 mixing chamber to beam center [mm]l61296 296 339 250 294 350
distance from M6 thread to beam center [mm]
adjustable
a 51+-10 51+-10 <=97 <=194249+-10<105
allowed size below beam center [mm]b 32 32 60 24** 65 59
(*) The maximum allowed size depends on the distance to the next spacer, which is larger when using the weak link. Please contact Markus Zolliker for designing sample holders larger than 14 mm in diameter.
(**) Slightly higher values might be acceptable on request.

On the mixing chamber as well as at the end of the sample holder/weak link there is an M6 hole, 6 mm deep. The (old) weak link has a length of m = 245 mm.


Attachments:  Dil2Samples.png 
This particular version was published on 28-Jun-2012 06:07:42 UTC by MarkusZolliker.