used in | | VARIOX | MA15 | MA11 | MA02 | MA09 | MA6 | MA7 |
sample space diameter [mm] | | 45 | 17 | 17 | 23.6 | 45 | 23.6 | 36 |
max. sample diameter [mm] | d | 40 | 14* | 14* | 22 | 40 | 22 | 32 |
distance from DIL1 mixing chamber to beam center [mm] | l | 83 | n.a. | n.a. | 339 | 272 | n.a. | n.a. |
distance from DIL2 mixing chamber to beam center [mm] | l | 83 | 296 | 296 | 339 | 272 | 294 | n.a. |
distance from DIL3 mixing chamber to beam center [mm] | l | 61 | 296 | 296 | 339 | 250 | 294 | 350 |
distance from DIL3 M6 thread to beam center [mm] adjustable | a | | 51+-10 | 51+-10 | <=97 | <=194 | 249+-10 | <105 |
allowed size below beam center [mm] | b | | 32 | 32 | 60 | 24** | 65 | 59 |
(*) The maximum allowed size depends on the distance to the
next spacer, which is larger when using the weak link. Please contact
On the mixing chamber as well as at the end of the
sample holder/weak link there is an M6 hole. The (old) weak link
has a length of m = 245 mm.