This is version 2.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
DIL1/DIL2/DIL3 Sample mounting
used in | | MA15 | MA11 | MA02 | MA09 |
sample space diameter [mm] | | 17 | 17 | 23.6 | 45 |
max. sample diameter [mm] | d | 14* | 14* | 22 | 40 |
distance from M6 thread to beam center [mm] adjustable | a | 51+-10 | 51+-10 | <=97 | <=194 |
allowed size below beam center [mm] | b | 32 | 32 | 60 | 24** |
(*) The maximum allowed size depends on the distance to the
next spacer, which is larger when using the weak link. Please contact
Markus Zolliker for designing sample holders larger than 14 mm in diameter.
(**) Slightly higher values might be acceptable on request.
On the mixing chamber as well as at the end of the
sample holder there is an M6 hole. The sample holder
has normally a length of m = 245 mm.
Attachments:
Dil2Samples.png
This particular version was published on 26-May-2011 15:18:10 UTC by MarkusZolliker.
|