DIL2/DIL3 Sample mounting

used in MA15MA11MA02MA09
sample space diameter [mm] 171723.645
max. sample diameter [mm]d14*14*2240
distance from M6 thread to beam center [mm]
adjustable
a51+-1051+-10<=97<=194
allowed size below beam center [mm]b32326024**
(*) The maximum allowed size depends on the distance to the next spacer, which is larger when using the weak link. Please contact Markus Zolliker for designing sample holders larger than 14 mm in diameter.
(**) Slightly higher values might be acceptable on request.

On the mixing chamber as well as at the end of the sample holder there is an M6 hole. The sample holder has normally a length of m = 245 mm.