Line 19 was replaced by lines 19-23 |
- ... |
+ * cool down the heat exchanger of the variox or magnet (setpoint 5 K) |
+ * pump the IVC (15 min. with sorb temperature set to 40 K, |
+ then 1 h without sorb heater) |
+ * test if the temperature of the mixing chamber is no longer coupled |
+ to the heat exchanger (changing the setpoint to 6 or 7 K). |